Hardware Functions
Measured Value
Judgement Outputs
HIGH output
PASS output
Relationship between Measured Value and Judgement Outputs
Linear Output
The linear output can be switched between current output and voltage out-
put.
Output current range: 4 to 20 mA (default value: 4 to 20 mA)
Output voltage range: −5 to 5 V (default value: −4 to + 4 V)
The output scale can be changed using the monitor focus function. The
maximum value of the current output is 23 mA and the maximum value of
the voltage output is 5.5 V.
Linear Output GND
The linear output GND is the ground for the linear output. Connect it sepa-
rately from the normal GND (0 V).
Always connect the linear output GND to ensure linearity and resolution for
the linear output.
36
HIGH threshold
Measured value
LOW threshold
ON
OFF
ON
OFF
ON
LOW output
OFF
Section 3-2
Hysteresis
Operating point
Releasing point