4.6.9 Code 22 : Maximum Preparation Time

START
ALLOW 2 TO 10 SEC TIME IN BETWEEN
FIRST & SECOND STEP OF RADIOGRAPHY
TAKE AN EXPOSURE
ERROR CONTINUES
?
Yes
NO
REPLACE D915 PCB
STOP