Service Instructions
4.5.3 Code 05 : kVist <> 0
START
X8 & X1
MAKE THEIR
KVIST = 0 ?
(ON D915 & D800) ARE
PROPER
No
No
(NO SPIKES)
CONNECTED
CONNECTION
PROPERLY ?
Yes
SWITCH OFF THE UNIT
Yes
R=20 Kohms AT
NO
X8.1,X8.2 & X8.3
REPLACE D800 PCB
YES
ON THE SINGLE
TANK.
REPLACE D915
STOP