Tighten the clamping knob.
Rotate the X-ray source aassembly so that the collimator points towards the floor.
Loosen the two Socket head Screws with a 8mm allen key three to four turns for
each screw.
Loosen the other four Socket head Screws with a 5mm allen key as shown in the
above figure.
Grab the handles and remove the X-ray source assembly. Place it on a soft bedding
with collimator facing upwards.
WARNING
3.2
Checking the line voltage, fuses and LEDs
3.2.1 Checking the line voltage
Measure the supply voltage at site using the digital Multimeter.
Ensure that the supply conditions
• Voltage
• Frequency
are within limits as specified in the Technical Specifications for the Unit.
Switch the Unit ON
Note : If the Unit cannot be switched ON
Check Supply at Mains socket.
Check Continuity of mains cable with the Plug pins.
Check Over Current Protective devices U1 & U2.
Switch the Unit OFF.
Siemens Ltd. Med Division India
Do not attach anything other than the X-ray source
assembly to the arm system.
Other objects will disengage the transport safety
lock,
causing
the
unanticipated which can cause both personal injury
and damage to the equipment.
Version 4.0
Service Manual
Service Instructions
arm
system
to
move
Page 18 of 48