4.6.6 Code 17 : Exposure terminated by Backup Timer

MULTIPLY JR & PULSE
DURATION OF J1.9
IS
IT EQUAL TO mAS
?
Yes
ERROR CONTINUES
?
Siemens Ltd. Med Division India
START
MONITOR JR ON D915 &
mAS ON D800 PCBS
ARE PULSES ?
YES
NO
ADJUST FREQUENCY AT J1.9 BY
No
NO
VARYING P1
TAKE AN EXPOSURE AND
NOTE THE TIME
TIME EXCEEDS
EXPOSURE TIME ?
Yes
Yes
REPLACE D915 PCB
STOP