4.6.5 Code 15 : JR < JRS
Check and
replace D800 if
found fauly.
Siemens Ltd. Med Division India
SWITCH OFF THE UNIT
CONFIRM mAS LINK ON
CHECK 'I' FOR 0.6V
+/ -0.1V AC RMS
Check for 22+/- 3 Ohms between
X8.5 and X8.6 and replace D800 if
Voltage betw. X8.5
No
and X8.6 is OK?
Yes
Check and replace
D915 if found faulty
START
D800 PCB
found faulty.
KV build up OK?
Yes
Version 4.0
Service Manual
Service Instructions
No
Check Inverter ckt/
Inverter pulses/
Resonance cap. C5.
STOP