4.6.3 Code 13 : Iheiz > Imax OR JR > Jrmax
DECREASE BY VARYING
POT P5 AND P6
MONITOR I FOR
PULSES
NO
REPLACE D801 PCB
Siemens Ltd. Med Division India
START
MONITOR IH AND JR ON
D915 PCB
IS
IH MAX ERROR
YES
?
YES
TAKE AN EXPOSURE
ERROR CONTINUES ?
STOP
NO
IS JR>JRmax ?
NO
REPLACE D915 PCB
No
Version 4.0
Service Manual
Service Instructions
YES
REDUCE OVERSHOOTS
IN JR BY PRESET P5
ELSE REPLACE D915
YES
Page 38 of 48