KVist < kVsoll
15
JR < JRS
17
Backup Timer
18
Premature Exposure
Termination
21
I > Iheiz
22
Maximum Preparation Time
Siemens Ltd. Med Division India
Maximum value of mAist above which Code is displayed is 180mA.
kVist is continuously polled during exposure. The value of kVist should be
greater than 85% of kVsoll
JR is continuously polled during exposure. The value of JR should be greater
than 50% of JRS.
If regular means of terminating exposure fails and exposure gets terminated
by the backup timer.
Exposure Release Switch is released before the exposure is terminated by
the mAs Integrator.
Same as Code 04, but during Exposure.
This Code is displayed if First step is pressed for more than 15 sec.