4.6.2 Code 12 : kVist > kVmax
START
MONITOR KVM & REG SIGNAL AND
TAKE AN EXPOSURE
KVM PULSE
Yes
IS REG>15KHz ?
Yes
SWITCH OFF THE UNIT AND SHORT
THE ST LINK ON D915 PCB
SWITCH ON THE UNIT & SELECT THE
Pr.7 BY PRESSING KV+
PRESS DL_SERV KEY AND THEN
EXPOSURE RELEASE SWITCH
MONITOR REG AND SET THE
FREQUENCY TO 15KHz. BY VARYING
POT P4
A
STOP
Siemens Ltd. Med Division India
ERROR CONTINUES
No
YES
NO
SWITCH OFF THE UNIT
X8.1,X8.2 W.R.T. X8.3 ON
Yes
REPLACE D915
ERROR CONTINUES
Version 4.0
Service Manual
Service Instructions
NO
?
A
IS
20K AT PIN
D915
PCB ?
REPLACE D800 PCB
ON SINGLE TANK
Yes
?
REPLACE SINGLE
No
No
TANK