3.2.2 Checking the fuses

Open the front cover of the control unit.
Loosen the Fuse carrier, which are mounted on the transformer bracket of the
control unit and check the continuity of fuse link. If fuse link has responded for
overcurrent replace the Fuse Link.

3.2.3 Checking the LEDs on D801

F1
6AT
F2
1AT
F3
2.5AT
F4
2.5AT
F5
10AT
F6
0.8 AT
F7
1.6AT

Standby mode

On D801 the following LED's are illuminated
V30
Supply for SMPS
V31
Supply for Filament Inverter.
V32
Supply for Rotating Anode ckt
V35
12V AC Supply for Collimator
V36
+VP supply
V50
24V DC on D801
V23&V24
DC supply for filament inverter
3.3

Checking the Control Voltages

In the standby state of the Unit,measure the voltages DC at X15 connector of D915
PCB.
Pin
Signal Name
No.
1
2
3
4
DC supply
5
DC supply
6
7
DC supply
- nc - = No Connection
Siemens Ltd. Med Division India
Filament Inverter supply.
SMPS supply
Rotating Anode supply .
24V Ac supply for Control ckt.
Collimator supply
DC-DC Conv. ckt.
220V AC supply for contactors.
Input/Output
Dgnd
-nc-
Dgnd
Agnd
max. permissible
voltage/current
Input
-nc-
Input
Input
Input
Input
Input
Version 4.0
Service Manual
Service Instructions
0V
-
0V
+ 5V
+15V
0V
-15V
Page 19 of 48

Hide quick links:

Advertisement

Table of Contents
loading