4.6.10 Code 10 : Rotating Anode not OK

Is ARR signal
obtained on D915?
No
Replace D915
Siemens Ltd. Med Division India
ALLOW 2 TO 10 SEC TIME IN BETWEEN
FIRST & SECOND STEP OF RADIOGRAPHY
Is anode in the
SingleTank rotating ?
No
220V AC supply to
Yes
Rot. Anode ckt OK?
Check fuse F3 on D801
and replace if found
If relay K2 on D801
doesn't operate,replace
Service Instructions
START
No
faulty.
K2
stop
Version 4.0
Service Manual
Yes
Check feedback
voltages on test points
I1 and I2.
Yes
Replace D801