Connecting The Replenishment Tanks To The Replenishment Pump; Installing The Silver Recovery Unit - Kodak X-OMAT M35 Installation Instructions Manual

Processors
Hide thumbs Also See for X-OMAT M35:
Table of Contents

Advertisement

Connecting the Replenishment Tanks to the Replenishment Pump

• Do not allow water in the tubing from
the REPLENISHMENT TANKS.
• The highest solution level in the
REPLENISHMENT TANKS must be below
the solution level in the processor TANKS.
• Maximum processor solution level is 97 cm
(38 in.) with the processor installed on an
M35 MOUNTING STAND.
[1]
Connect the SILVER RECOVERY UNIT to the
fixer DRAIN with ⁄
18

Installing the Silver Recovery Unit

1
-inch (1.27 cm) ID tubing.
2
[1]
Check that the connections in the tubing are
tight.
[2]
Install the 2 STRAINERS in the tubing
between the REPLENISHMENT PUMP and
the REPLENISHMENT TANKS. Use the ⁄
(9.5 cm) TUBING supplied.
[3]
Connect the tubing to the processor TANKS.
[4]
Check that the REPLENISHMENT TANKS are
connected to the correct processor TANKS.
REPLENISHMENT TANKS
STRAINERS
FLOOR DRAIN
OUT
SILVER RECOVERY UNIT
3
in.
8
H104_0026BCB
IN
H112_0118AC
981159