Calibrating the Replenishment System

Description:
Calibrate the replenishment system every 3 months.
Calibrating the replenishment system determines the actual rate of the processing solution flowing
through the Replenishment Pumps. To calibrate the Pump you need to first measure the volume of
solution pumped during a set time period, and secondly enter the information into the microprocessor
by using the procedure outlined below. The microprocessor then computes the flow rate of the solution
through the Pump and adjusts the length of time that the Pump must operate so that the volume of
replenishment delivered equals the replenishment volume selected.
Important
The volume of replenishment solution actually measured during this procedure is not the volume
delivered for a 35 x 43 cm sheet of film.
Procedure:
[1] From the Walk-Up Menu, press the "GO TO SETUP" key.
READY
[2] Enter the 4-digit access code.
[3] Press the "MORE" key.
[4] Press the "SETUP" key.
[5] Press the "PUMP CALIB" key.
PROCESS
[6] Press the "DEV CAL" key to calibrate the Developer Pump or the "FIX CAL" key to calibrate the Fixer Pump.
30
DRYER
SLEEP
TEMP
1
2
INFO
SETUP
CLOCK
DEV
FIX
CAL
CAL
SELECT
MORE
CYCLE
3
4
CYCLE
MORE
OPTIONS
AUTO
PUMP
STARTUP
CALIB
STD
GO TO
SETUP
CANCEL
REQUEST
DONE/
RETURN
DONE/
RETURN
DONE/
RETURN
DONE/
RETURN
30SEP98 – 5B6328

Advertisement

Table of Contents
loading

Table of Contents