Table 3-1: Analog Concentration Output-Example; Table 3-2: Analog Range Id Output-Example - Teledyne 3000TB Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3000TB:
Table of Contents

Advertisement

Installation
Examples:
The analog output signal has a voltage which depends on the
oxygen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the autoranging mode.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on a range that was
defined as 0–100 ppm O
Table 3-1: Analog Concentration Output—Example
ppm O
2
0
10
20
30
40
50
60
70
80
90
100
To provide an indication of the range, a second pair of analog
output terminals are used. They generate a steady preset voltage (or
current when using the current outputs) to represent a particular range.
Table 3-2 gives the range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
LO
MED
HI
CAL (0-25%)
Teledyne Analytical Instruments
, then the output would be as shown in Table 3-1.
2
Voltage Signal
Output (V dc)
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Voltage (V)
0.25
0.50
0.75
1.00
Model 3000TB
Current Signal
Output (mA dc)
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
Current (mA)
8
12
16
20
24

Advertisement

Table of Contents
loading

Table of Contents