Trace Oxygen Analyzer

3.3.3 Analog Outputs

There are eight DC output signal connectors on the ANALOG
OUTPUTS connector block. There are two connectors per output with
the polarity noted. See Figure 3-5.

Figure 3-5: Analog Output Connections

The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing oxygen,
0–1 V dc Range ID:
4–20 mA dc % Range: Current increases linearly with increasing
4–20 mA dc Range ID: 8 mA = Low Range
Teledyne Analytical Instruments
from 0 V at 0 ppm to 1 V at full scale ppm.
(Full scale = 100% of programmable range.)
0.25 V = Low Range
0.5 V = Medium Range
0.75 V = High Range
1 V = Air Cal Range
oxygen, from 4 mA at 0 ppm to 20 mA at full
scale ppm. (Full scale = 100% of programmable
range.)
12 mA = Medium Range
16 mA = High Range
20 mA = Air Cal Range
Installation
23