Temperature Calibration Procedure

Tools required
Temperature measuring device (see note below)
Parts required
#
1
For the measuring and calibration process Agilent Technologies recommends a
N O T E
thermometer with 0.1 °C precision. Contact the local Agilent Technologies support
representative for ordering information.
The figures in this procedure refer to a specific type of temperature sensor (Heraeus
N O T E
Quat340, quartz surface-temperature measurement sensor). Other sensors may require a
different fixing.
1 Install the temperature sensor (Installing the Temperature Sensor)
Temperature Calibration with Agilent LabAdvisor
This is the standard 1-point calibration. If a 2-point calibration is required,
refer to
If only one sensor is available, the procedure has to be performed for both, the
left and right heat exchanger separately.
Agilent 1260 Infinity TCC User Manual
Description
Calibrated temperature measuring device
("Installing the Temperature Sensor"
"Two Point Calibration"
Column Thermostat Temperature Calibration
on page 90).
on page 86
7
Test Functions
83