Column Thermostat Temperature Calibration Procedure - Agilent Technologies 1200 Series User Manual

Thermostatted column compartment
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

6
Test Functions
Column Thermostat Temperature Calibration

Column Thermostat Temperature Calibration Procedure

Tools required
Temperature measuring device (see note below)
Parts required
Description
Calibrated temperature measuring device
For the measuring and calibration process Agilent Technologies recommends a
N O T E
thermometer with 0.1 °C precision. Contact the local Agilent Technologies support
representative for ordering information.
The figures in this procedure refer to a specific type of temperature sensor (Heraeus,
N O T E
Quat340, quartz surface-temperature measurement sensor). Other sensors may require a
different fixing.
1 Install the temperature sensor
2 Select the column-compartment temperature calibration mode in the user
3 Wait for the temperature to stabilize at the calibration temperature (36 °C).
4 Measure the temperature of the heat exchanger.
5 If the measured temperature deviates by more than ± 0.5 °C from the actual
6 Install the sensor at the measurement point on the right heat exchanger.
Limits
N O T E
After calibration, the measured temperature and the calibration temperature should be
within ± 0.5 °C. The maximum deviation which can be adjusted is ± 1.6 °C. If the measured
value and the calibration value differ by more than ± 1.6 °C, this is an indication that a
problem exists,
68
page 69).
interface.
temperature, enter the measured value in the measured-temperature field
for the left heat exchanger.
Repeat the calibration procedure for the right heat exchanger.
"Column Thermostat Calibration Problems"
("Installing the Temperature Sensor"
on page 69.
1200 Series TCC User Manual