PS200
PORTABLE GAS MONITOR
MANUAL BUMP TEST OPTIONS
Figure 6-4: Bump Test Kit

6.3. Quick/Full Bump Test

Operation now depends on whether QUICK or FULL bump is configured.
6.3.1. QUICK BUMP TEST
As the alarm threshold for each range is exceeded, the audible/visual/vibration alarms will activate
and a
symbol will appear, otherwise a
symbol will be displayed.
6.3.2. FULL BUMP TEST
After a short period of time, the gas readings are checked against configurable limits. The audible/
visual/vibration alarms will activate and a
symbol will appear if the readings are within these
limits, otherwise a
symbol will be displayed.

6.4. Confirm Alarms

The user is prompted to confirm if the audible and visual alarms were activated, as shown in
Figure 6-5: Confirm
Alarms.
Figure 6-5: Confirm Alarms