PS200
PORTABLE GAS MONITOR
CALIBRATION
NOTE: The test gas must match the predefined cylinder values stored in the monitor.
5. Once gas is applied, press and hold the R button to start the calibration. The 60 second count-
down will commence (refer to
will be calibrated to the predefined values.
6. The timer will count up to 9 seconds, during which time the monitor will adjust the gain to the
sensors accordingly. No manual input is required.
7. Once calibration is successfully completed, the user has the option of saving the results (see
Figure 9-6: CAL COMPLETE/SAVE RESULTS
Figure 9-6: CAL COMPLETE/SAVE RESULTS Screen
8. Using the L button, highlight one of the 3 the available options:
64171, Rev. 11/December 2019
Figure 9-4: APPLY GAS
Figure 9-4: APPLY GAS Screen
Figure 9-5: APPLY GAS Screen
Screen).
Screen). After 60 seconds, the monitor
9-3