NOTE: Ensure that the locating pegs in the monitor filter recess align with the mating
pin-holes in the hydrophobic filter.
Place the filter cover over the filter recess and slide it towards the display until the locating lugs
engage in the locating slots.
Using a No.1 Pozidrive
NOTE: Do not overtighten the captive screw.
8.2.2. SAMPLE INLET (DUST) FILTER REPLACEMENT
Using a No.1 Pozidrive
Remove the inlet nozzle. The sample inlet filter is located at the rear of the inlet nozzle.
Push a matchstick, or similar, into the front of the inlet nozzle and remove the sample inlet filter
from the rear of the inlet nozzle.
Insert a new sample inlet filter into the rear of the inlet nozzle. Ensure the 'rough' surface faces
the inlet filter (sample side).
Replace the inlet nozzle. The inlet nozzle will only fit in one direction.
Using a No.1 Pozidrive
NOTE: Do not overtighten the screws.
8-2
screwdriver, tighten the captive screw.
®
Figure 8-2: Sample Inlet (Dust) Filter Replacement
screwdriver, remove the 2 Pozi Pan screws.
®
screwdriver, insert the 2 Pozi Pan screws.
®
PS200
PORTABLE GAS MONITOR
OPERATOR MAINTENANCE
64171, Rev. 11/December 2019