6.4.1. BUMP TEST – PASS
After selecting YES, BUMP TEST PASS is displayed, as shown in
The bump test result including date and time is automatically datalogged.
When the gas readings fall below their alarm set-points, or after 60 seconds, the bump test is com-
plete, and the monitor will automatically return to normal operation.
6.4.2. BUMP TEST – FAIL
If any range fails the bump test, the CONFIRM ALARMS display will be red and a
displayed, as shown in
After selecting YES or NO, BUMP TEST FAIL is displayed and the user is instructed to switch the
monitor off, as shown in
If a monitor fails a bump test, then a full re-calibration is necessary.
6-4
Figure 6-6: Bump Test Pass
Figure 6-7: Confirm Alarms -
Figure 6-7: Confirm Alarms - Fail
Figure 6-8: Bump Test Fail - Switch Monitor
Figure 6-8: Bump Test Fail - Switch Monitor OFF
MANUAL BUMP TEST OPTIONS
Figure 6-6: Bump Test
Fail.
OFF.