Agilent Technologies N5230AU/CU-971 Installation Note page 34

Extended frequency range upgrade kit to upgrade option 220 to option 420
Table of Contents

Advertisement

Step 16. Perform Post-Upgrade Adjustments
Adjustments
The following adjustments must be made due to the hardware changes of the analyzer.
• default EE adjustment - select all assemblies, one at a time, and initialize.
• source calibration
• receiver calibration
These adjustments are described in the PNA service guide and in the PNA on-line HELP. A
list of equipment required to perform these adjustments can be found at
Required for Post-Upgrade Adjustments" on page
Calibration
The analyzer should now operate and phase lock over its entire frequency range.
If you experience difficulty with the basic functioning of the analyzer, contact Agilent. Refer to
"Getting Assistance from Agilent" on page
• Although the analyzer functions, its performance relative to its specifications has not been
verified.
• It is recommended that a full instrument calibration be performed using the PNA
performance test software.
• Refer to the analyzer's service guide for information on the performance test software.