Agilent Technologies N5230AU/CU-971 Installation Note page 23

Extended frequency range upgrade kit to upgrade option 220 to option 420
Table of Contents

Advertisement

Step 9. Install the 50 GHz Top to Bottom Cables
Follow instructions carefully when making cable connections, especially
CAUTION
wire harness connections. Incorrect connections can destroy components,
resulting in additional customer costs.
1. Place the analyzer on its side as shown in
page
7.
2. Insert cables W3 and W6 into the analyzer in the direction indicated by the arrows. It may
be necessary to guide the other end of the cables, on the top side of the analyzer, as they
are being inserted.
3. Loosely connect cables W3 and W6 at the locations shown in
all connectors are loosely connected, tighten the connectors.
Installation Note N5230-90010
Figure
13. New parts are listed in
Figure 13
Table 1 on
and
Figure
14. Once
23