Step 4. Remove the 20 GHz Bottom Cables
Refer to either
Figure 5
prefix of your analyzer.
1. Place the analyzer bottom-side up on a flat surface.
2. Remove the following cables in the order listed:
• W8
A17 MASS 26.5 to A21 test port 1 coupler
• W10 A17 MASS 26.5 to A20 mixer brick (R1)
• W12 A21 test port 1 coupler to A20 mixer brick (A)
• W9
A17 MASS 26.5 to A22 test port 2 coupler
• W11 A17 MASS 26.5 to A20 mixer brick (R2)
• W13 A22 test port 2 coupler to A20 mixer brick (B)
Figure 5
20 GHz Bottom Cables Removal (serial numbers prefixed MY4500 and below)
Installation Note N5230-90010
or
Figure 6
for this procedure, depending upon the serial number
13