Ac Adapter And Battery; Charge The Battery - Fluke CableAnalyzer DSX-602 User Manual

Table of Contents

Advertisement

DSX-602 CableAnalyzer
Users Manual

AC Adapter and Battery

You can use the AC adapter (model PWR-SPLY-30W) or the
lithium ion battery (model VERSIV-BATTERY) to supply power to
the tester.
To remove the battery, see "Remove the Battery" on page 100.

Charge the Battery

Before you use the battery for the first time, charge the battery
for about 2 hours with the tester turned off.
To charge the battery
Connect the AC adapter to the 15V jack on the left side of the
tester. The LED near the AC adapter connector is red when the
battery charges, and green when the battery is fully charged.
A fully-charged battery operates for approximately 8 hours of
typical use. The battery takes approximately 4 hours to fully
charge when the tester is turned off.
You do not need to fully discharge the battery
before you recharge it.
The battery will not charge if its temperature is
outside the range of 32 °F to 104 °F (0 °C to 40 °C).
The LED near the connection for the AC adapter is
yellow if the battery will not charge.
14