Input Subsystem - Fluke Norma 4000 User Manual

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

Remote Control
Users Guide
INITiate[:IMMediate]:SEQuence1
INITiate[:IMMediate]:NAME STARt
Description
Initiates the memory acquisition start trigger. After initiation the pretrigger gets filled (if
> 0). Pretrigger is filled when "waiting for trigger" bit of OPER:STAT register is set to 1.
An alias for SEQuence1 is STARt.
Parameter(s)
-
*RST state
-
Example(s)
INITiate:NAME STARt
Invalidates
-
Invalidated by
-

INPut Subsystem

The INPut subsystem controls the characteristics of the input channels. Numeric suffixes
of the INPut node correspond to hardware input channel of the instrument. For 6 channel
models, the valid electrical channel suffixes are 1 to 6. For 12 channel models, the valid
electrical channel suffixes are 1 to 12. The electrical INPut subsystem does not
distinguish between current and voltage channels. Input filter settings are common for all
electrical channels, so the channel suffixes can be omitted, even if implemented for
compatibility reasons. If the channel suffix is omitted, the command applies to input 1.
For Process Interface option, the valid mechanical channel suffixes are 21 to 24 (torque
input 1..4) and 25 to 28 (speed input 1..4). See section 2.1.10 SENSe2 Subsystem for
detailed description.
Command
:INPut[1..12]
:COUPling
:GAIN
:FILTer
[:STATe]
[:LPASs]
:FREQuency?
:SHUNt
:INPut[21..28]
:TYPE
4-32
Parameter
AC | DC
0 to 1.0e12
ON | OFF
INTernal | EXTernal
VOLTage | FREQuency
Default Value/Unit
DC
1.0
Current channels only
ON
Device dependent
Query only
INTernal
Current channels only
Option Process Interface
Remark