Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

SENSe2:TORQue[1..4]:VOLTage:OFFSet:IMMediate
Description
Configure the currently measured torque voltage to be the offset value. The measurement
must be valid (no overload).
Parameter(s)
-
Example(s)
SENS2:TORQ3:VOLT:OFFS:IMM
*RST state
-
Invalidates
SENSe2:TORQue[1..4]:VOLTage:OFFSet
Invalidated by
-
SENSe2:TORQue[1..4]:FREQuency:SCALe <value>
Description
Sets the torque scaling factor for frequency type input that reflects conversion ratio of
torque sensors employed. The difference between the frequency on the respective input
and the specified offset is multiplied by this scaling factor before any further processing.
Parameter(s)
-1.0e6 to 1.0e6
[rpm/Hz]
Example(s)
SENS2:TORQ2:FREQ:SCAL 0.001
SENS2:TORQ1:FREQ:SCAL?
*RST state
1.0
Invalidates
-
Invalidated by
-
SENSe2:TORQue[1..4]:FREQuency:OFFSet[:VALue] <value>
Description
Sets the input frequency quantity that corresponds to torque zero values. This frequency
is subtracted from the measured frequency at the input before this difference is multiplied
by the scaling factor.
Remote Control - Description of Commands
Response: 0.001
Commands and Queries
4
4-55