Fluke Norma 4000 User Manual page 154

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

Remote Control
Users Guide
Error Number
-100
-101
-102
-103
-104
-108
-109
-110
-112
-113
-114
-120
5-4
ErrorDescription [description/explanation/examples]
Command error
[This is the generic syntax error for devices that cannot detect more specific errors. This
code indicates only that a Command Error as defined in IEEE 488.2, 11.5.1.1.4 has
occurred.]
Invalid character
[A syntactic element contains a character that is invalid for that type; for example, a
header containing an ampersand, SETUP&.]
Syntax error
[An unrecognized command or data type was encountered; for example, a string was
received when the device does not accept strings.]
Invalid separator
[The parser was expecting a separator and encountered an illegal character; for
example, the semicolon was omitted after a program message unit, *SRE 1:INP1:COUP
AC.]
Data type error
[The parser recognized a data element different than one allowed; for example, numeric
or string data was expected but block data was encountered.]
Parameter not allowed
[More parameters were received than expected for the header; for example, the *SRE
common command only accepts one parameter, so receiving *SRE 2,1 is not allowed.]
Missing parameter
[Fewer parameters were received than required for the header; for example, the *SRE
common command requires one parameter, so receiving *SRE is not allowed.]
Command header error
[An error was detected in the header.]
Program mnemonic too long
[The header contains more than twelve characters (see IEEE 488.2, 7.6.1.4.1).]
Undefined header
[The header is syntactically correct, but it is undefined for this specific device; for
example, *XYZ is not defined for any device.]
Header suffix out of range
[The value of a numeric suffix attached to a program mnemonic, see Syntax and Style
section, makes the header invalid.]
Numeric data error
[This error is generated when parsing a data element that appears to be numeric,
including the nondecimal numeric types. For example INP:GAIN 1.0X2 will generate this
error.]