Device-Specific Error - Fluke Norma 4000 User Manual

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

Device-Specific Error

An <error/event number> in the range [ -399 , -300 ] or [ 1 , 32767 ] indicates that the
instrument has detected an error, possibly due to an abnormal hardware or firmware
condition. These codes are also used for self-test response errors. The occurrence of any
error in this class causes the device-specific error bit (bit 3) in the event status register
(IEEE 488.2, section 11.5.1) to be set.
Error Number
-300
Device-specific error
[This is the generic device-dependent error for devices that cannot detect more specific
errors. This code indicates only that a Device-Dependent Error as defined in IEEE
488.2, 11.5.1.1.6 has occurred.]
-310
System error
[Indicates that some error, termed "system error" by the device, has occurred. This code
is device-dependent.]
-311
Memory error
[Indicates some physical fault in the device's memory, such as parity error.]
-313
Calibration memory lost
[Indicates that nonvolatile calibration data used by the *CAL? command has been lost.]
-314
Save/recall memory lost
[Indicates that the nonvolatile data saved by the *SAV? command has been lost.]
-315
Configuration memory lost
[Indicates that nonvolatile configuration data saved by the device has been lost.]
-320
Storage fault
[Indicates that the firmware detected a fault when using data storage. This error is not an
indication of physical damage or failure of any mass storage element.]
-325
Sample factor adjusted
[Indicates that applying current configuration caused the sample factor to be adjusted.]
-326
Recording time too long
[Indicates that the data acquired within the specified recording time would not fit into the
available memory.]
-330
Self-test failed
-340
Calibration failed
-350
Queue overflow
[A specific code entered into the queue in lieu of the code that caused the error. This
code indicates that there is no room in the queue and an error occurred but was not
recorded.]
-360
Communication error
Error String [description/explanation/examples]
5
Error Messages
Device-Specific Error
5-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Norma 5000