Parameter(s)
bit 2
Ranging (changing range).
bit 3
Sweeping (memory recording in progress).
bit 5
Waiting for trigger.
bit 8
Synchronized (if sync source changes, there is a glitch).
bit 10
Averaging (averaging in progress, at the end of each averaging cycle, there is a glitch).
bit 12
Spectrum CALCulation in progress.
*RST state
0
Example(s)
STAT:OPER:NTR 5948
STAT:OPER:NTR? Response: 5948
Invalidates
-
Invalidated by
-
STATus:OPERation[:EVENt]?
Description
This query returns the contents of the event register associated with the status structure
defined in the command. The response is (NR1 NUMERIC RESPONSE DATA) (range:
0 through 32767) . Note that reading the event register clears it.
Response
<value> is a 16 bit integer number in decimal notation.
bit 2
Ranging (changing range).
bit 3
Sweeping (memory recording in progress).
bit 5
Waiting for trigger.
bit 8
Synchronized (if sync source changes, there is a glitch).
bit 10
Averaging (averaging in progress, at the end of each averaging cycle, there is a glitch).
bit 12
Spectrum CALCulation in progress.
*RST state
Has no effect.
Example(s)
STAT:OPER?
Invalidates
-
Invalidated by
-
Response: 2 (changing range)