Fluke Norma 4000 User Manual page 100

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

Remote Control
Users Guide
Input
Channel Suffix
INPut21 (torque)
INPut25 (speed)
INPut22 (torque)
INPut26 (speed)
INPut23 (torque)
INPut27 (speed)
INPut24 (torque)
INPut28 (speed)
Command
:INPut[21..28]
:TYPe
INPut[21..28]:TYPe VOLTage | FREQuency
Description
Selects the type of signal measured for a Process Interface input.
Parameter(s)
VOLTage
FREQuency
*RST state
FREQuency for all Process Interface inputs
Example(s)
INP21:TYP VOLT
INP25:TYP?
Invalidates
-
Invalidated by
-
The SENSe2 node distinguishes the mechanical from the electrical system (SENSe1 for
the electrical system is the default node of the root level of the command tree). The
SENSe2:xxx:VOLTage nodes apply if the corresponding input type is set to
INPutx:TYPe VOLTage and SENSe2:xxx:FREQuency nodes apply if the corresponding
input type is set to INPutx:TYPe FREQuency respectively.
4-52
Drive
Index
1
SENSe2:TORQue|SPEed|POLepairs|TYPe|REFerence1[:POWer]
2
SENSe2:TORQue|SPEed|POLepairs|TYPe|REFerence2[:POWer]
3
SENSe2:TORQue|SPEed|POLepairs|TYPe|REFerence3[:POWer]
4
SENSe2:TORQue|SPEed|POLepairs|TYPe|REFerence4[:POWer]
Parameter
VOLTage | FREQuency
Input signal is voltage.
Input signal is frequency.
Response: FREQ
Sense Subsystem Phase Suffix
Default
Value/Unit
FREQuency
Analog/digital sensor
Remark