Fluke Norma 4000 User Manual page 125

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

TRIGger:STARt:LEVel <level>
Description
When the start source for recording is an averaged measurement function, then this
setting specifies the measurement function level that will trigger the recording.
Parameter(s) <level>
Range for this setting is not defined.
*RST state
0.0
Example(s)
TRIG:STARt:LEV 50.0
TRIG:STARt:LEV? Response: 25.0
Invalidates
-
Invalidated by
-
TRIGger:STARt:SLOPe POSitive | NEGative
Description
When the stop source for recording is an averaged measurement function, then this
setting specifies the slope.
Parameter(s)
POSitive
Triggers on positive slope.
NEGative
Triggers on negative slope.
*RST state
POS
Example(s)
TRIG:STAR:SLOP POS
TRIG:STAR:SLOP?
Invalidates
-
Invalidated by
-
TRIGger:STOP:SOURce TIME | IMMediate | MANual | <function>
Description
Stop trigger source. The acquisition will stop either at specified date/time, when memory
is full, recording time is reached, or the selected stop condition below is met, whichever
comes first.
Response: POS
Remote Control - Description of Commands
Commands and Queries
4
4-77