Fluke Norma 4000 User Manual page 65

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

For FFT frequency step is identical for all functions, for DFT every function can have an
individual fundamental frequency. If no fundamental frequency can be found, an ASCII
equivalent of NaN is returned (+9.91E+37).
*RST state
Example(s)
CALC:DATA:PRE?
Invalidates
-
Invalidated by
-
CALCulate:INTegral[:STATe] ON | OFF
Description
Controls the state of integration capability of the instrument. When enabled, the
instrument can integrate on some averaged measurement functions.
Parameter(s)
ON
Integration enabled.
OFF
Intergration disabled.
Example(s)
CALC:INT ON
CALC:INT?
*RST state
OFF
Invalidates
-
Invalidated by
-
CALCulate:INTegral:FUNCtion <function>{,<function>}
Description
Specifies function list for integral calculation. The <function> is specified as a quoted
string. For example: "POWer1".
Comma separated list of <function> may be sent as parameters. Applying a new list
invalidates the previously active list (if any).
The query response returns a comma separated list of functions, each of which are
<STRING RESPONSE DATA>, i.e. enclosed in quotes. The query returns the short form
mnemonics and omits any default nodes in the <function>.
Averaged POWer and VOLTage/CURRent:MEAN values can be integrated. Specifiers
INTegral | PINTegral | NINTegral can be used then in the SENSe:FUNCtion /
SENSe:DATA? list to read the integrated values. If the integration function list is
Response: 11.32, 40, 3, 50.0,50.0,50.0
Response: ON
Remote Control - Description of Commands
Commands and Queries
4
4-17