A1 VCXO Frequency Adjustment

The purpose of this procedure is to adjust the VCXO on the A1 Reference Oscillator
Board.

Required Equipment

Frequency Counter
BNC(m)-BNC(m) Cable, 61 cm

Procedure

Step 1. Run the adjustment program.
Step 2. Choose the A1 Adjustment in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Make sure that nothing is connected to the "Ext Ref In" on the E4991A rear panel.
Step 4. Connect the instrument as shown in Figure 3-10
Figure 3-10
A1 VCXO Frequency Adjustment Setup
Step 5. Follow the adjustment program instruction to update the correction constant.
Chapter 3
Description
53181A Opt.010 and 030,
53131A Opt.010 and 030 or
53132A Opt.010 and 030
p/n 8120-1839
A1 VCXO Frequency Adjustment
Recommended Model
Adjustment
133