8 .4

Compatible X-ray system specification

IMPORTANT
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• AeroDR Detector is allowed to be connected to an X-ray apparatus of 10.3 sec. in its maximum irradiation time.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
1. Electrical specifications (Hardware requirement)
 XIF Board
Electro-mechanical relays are used for the interface.
The power for the relays needs to be supplied by the generator through the signal lines.
(1) Input Interface
Followings are the coil ratings of the relays for the
available).
Coil rated voltage
100[V]
120[V]
AC
220[V]
50/60[Hz]
230[V]
240[V]
DC
24[V]
(2) Output Interface
Followings are the contact ratings of the
Load type
Rated load
Resistance
AC 250[V]/2[A]
load
DC 30 [V]/2[A]
Inductive
AC 250[V]/0.8[A]
load
DC 30 [V]/1.2[A]
 XGIF Board
(1) "XRAY-READY", "XRAY-START", and "XRAY-ON" input units
Photocoupler ratings for "XRAY-READY", "XRAY-START", and "XRAY-ON" input units are as follows.
Only use 12 V ISO for sending or receiving signals from the X-ray generator.
Item
1
Signal-ON current*
1
Signal-OFF dark current*
Response cycle
12 V ISO output voltage
12 V ISO output current*
2
12 V ISO overload protection
The specifications for the "KM READY" output part are as follows.
The "KM READY" output part is a relay or photocoupler depending on the connected X-ray device.
When the "KM READY" output part is a relay
Load type
AC 250[V]/4[A]
Resistance load
DC 30 [V]/4[A]
In addition,regardless of the above specifications, the rated current is limited to 1 [A] when the interface cables pro-
vided by Konica Minolta, Inc. are used.
Power
Minimum operation
consumption
voltage
1.5[VA]
80[%] (min)
0.8[W]
"KM_Ready" output.
Rate
Contact voltage
current
AC 300[V] (max)
2[A]
DC 100[V] (max)
MIN
15
11
Rated load
Maximum load
AC 250[V]/4[A]
DC 110 [V]/0.2[A]
"Prepare", "Exposure" and "MainGate" inputs (six different voltages are
Maximum return
voltage
10[%] (max)
Contact
Opening/closing
current
capacity
AC 500[VA]
DC 60[W]
2[A] (max)
AC 200[VA]
DC 36[W]
TYP
10
Fuse (To recover, the board needs to be replaced.)
Contact current
AC 100[VA]
DC 120[W]
116
Maximum allowable
Opening/closing
voltage
3[sec/cycle] (min)
115[%] (max)
Minimum load
Resistance when
current
DC 5[V]/20[mA] 100[mΩ] (max)
MAX
10
13
60
*1: Value per circuit
*2: Total value for all input circuits
Minimum load
Resistance when
current
DC 1[V]/1[mA]
50[mohm] /(max)
frequency
closed
Unit
mA
uA
msec
V
mA