8.2 Pediatric patients' use;
<Individual patient Protection>
• Specific pre-set pediatric control settings that are
appropriate for the intended patient
• Exposure based on thickness, not age or weight-
based
(Exceptions such as portable neonatal chest)
• Configurable based on local preferences
<Grid Use>
• Grid double or triple exposure factors / patient
dose
• Grid remove scatter & improve image quality in
thicker patient
• Grid are not needed in thinner patients <10-12
cm of water thickness
• Easy to override default programming & not re-
move grid
<Added filtration>
• Removes low energy x-rays that don't contribute
to image formation but contribute to patient dose
• Little use of added filtration
<Electronic Cropping>
Image should be clearly marked and used for QA
review
<Hardware requirement>
• Improved Automatic Exposure Control (AEC)
design
(ex. Chamber designed for pediatric use)
• Dose Area Product (DAP) meters
Keeping DAP results report as part of a QA pro-
cess
<Software requirement>
• Strongly recommended to use Exposure Index
& Deviation Index comply with IEC 62494-1(As
stated in the beginning of this section).
<Exportable QA tool>
Image review and record to measure and compare
exposure and other parameters for each exam
performed.
These feature can be used as part of a facilities
quality program for education and continuous im-
provement to aim at AS LOW AS Possible expo-
sure.
<Education>
Please reflect updated terminology standard and
dose reporting profiles under medical imaging
physics or radiologist or radiologic technologist
representative.
CAUTION
Use of equipment and exposure settings designed
for adults of average size can result in excessive
radiation exposure for a smaller patient.
Studies have shown that pediatric patients may be
more radiosensitive than adults
(i.e., the cancer risk per unit dose of ionizing ra-
diation is higher), and so unnecessary
radiation exposure is of particular concern for pe-
diatric patients.
Hint
To keep following IEC 62494-1 correctly for using
our detectors, it is important to follow QC control
and compatible X-ray specification with our detec-
tors.
• Quality tool of AeroDR SYSTEM / AeroDR
SYTEM2 is prepared phantom and software
to assure detector within specification to be
assured before finding Exposure Index in ac-
cordance with the standard of Image Gen-
tly and government guideline as needed.
Please refer to "Chapter 6" of AeroDR SYSTEM
/ AeroDR SYSTEM 2 Operation manual.
• For further X-ray system specification to use with
AeroDR SYSTEM / AeroDR SYSTEM 2, please
refer to "8.4 Compatible X-ray system specifica-
tion".
112