Konica Minolta AeroDR System Operation Manual page 58

Direct digitizer
Hide thumbs Also See for AeroDR System:
Table of Contents

Advertisement

3.2 Operation of AeroDR Detector
 Exposure with the remote table
• The AeroDR Detector is precision equipment,
therefore impact or vibration during exposure and
image transfer may affect the image quality. Im-
pact or vibration when performing exposure using
the remote table in particular may significantly in-
crease the effect, therefore observe the following.
– Prevent the AeroDR Detector from moving un-
til the preview image is displayed immediately
after exposure.
– Do not operate the remote table to eject the
AeroDR Detector or load the AeroDR Detector
to the standby position.
– Do not move the remote table position horizon-
tally or vertically.
• Check that the AeroDR Detector does not auto-
matically move immediately after exposure. If it
does, please contact a technical representative
of the X-ray device manufacturer and request to
change the settings so that the AeroDR Detector
does not move immediately after exposure.
• If the settings of the X-ray device cannot be
changed, check the image quality in advance.
Please contact Konica Minolta technical repre-
sentative for more information about how to check
the image quality.
58

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aerodr system 2

Table of Contents