Communication Settings For Rs-232C - Agilent Technologies 1260 Infinity User Manual

Standard autosampler
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

1
Introduction to the Autosampler
Setting the 8-bit Configuration Switch (On-Board LAN)

Communication Settings for RS-232C

The communication protocol used in the column compartment supports only
hardware handshake (CTS/RTR).
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the column instrument
must be powered up again in order to store the values in the non-volatile
memory.
Table 8
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
Table 9
Table 10
34
Communication Settings for RS-232C Communication (without on-board LAN)
Mode
1
Select
RS-232C
0
Baudrate Settings (without on-board LAN)
Switches
3
4
5
0
0
0
0
0
1
0
1
0
0
1
1
Data Bit Settings (without on-board LAN)
Switch 6
0
1
2
3
4
1
Baudrate
Baud Rate
9600
1200
2400
4800
Data Word Size
7 Bit Communication
8 Bit Communication
Agilent 1260 Infinity Standard Autosampler User Manual
5
6
Data Bits
Switches
3
4
5
1
0
0
1
0
1
1
1
0
1
1
1
7
8
Parity
Baud Rate
9600
14400
19200
38400