Lifter column type A
electrically adjustable
Installing on type B lifter column - part 1
4 mm Allen wrench
Side cutting pliers
Cable ties
The conduits leading from battery box to seat must be
installed so they do not get crimped, stretched or sheared
when raising or lowering seat.
1. Run cables from actuator (2) of seat angle adjustment
to cable holder (1).
2. Secure cables with cable tie.
3. Run conduits further to lateral bar.
4.
Secure cables with cable ties (1) under outer side of
bar (2).
5.
1533060-M
Lifter column type B
electrically adjustable
View of seat from below
with ACT (1).
6.
Attach cable guide for lifter (1) using two screws, on
left underneath seat.
7. Position protective tube for cables completely from
outer side.
8. Secure protective tube with a cable tie (1) on cable
guide.
9. Proceed with "Installing on type B lifter column - part 2".
Installing on type B lifter column - part 2
1.
Run cable bundle, including protective tube for cables,
through opening in front shroud (2).
2. Secure cable bundle with cable tie under shroud.
3. Raise and lower seat several times while closely watching
cable bundle. The cable bundle must not get trapped,
stretched or crushed. The cable bundle must move
sideways, as shown in the following illustrations. The
drive motor must not be touched by cable bundle.
Lifter column completely
extended.
Service
View from rear.
Lifter column completely
retracted. Pay attention
to areas where cable
bundle might get crushed
or touch motor.