8. Command and Response Structure
8. Command and Response Structure
8. Command and Response Structure
8. Command and Response Structure
8. Command and Response Structure
4. Read alarm status character from the meter number 15 HEX:
Command format
Response format
where "@" is the alarm status character, decoded using Table 8.4.
5. Enable setpoints number 3 and 4 in point-to-point mode:
Command format
Response format
6. Hold display value on meter number 15 HEX:
Command format
Response format
7. Cold reset (reset 1) in point-to-point mode:
Command format
Response format
8. Publish measurement string from the meter in point-to-point mode:
Command format
Response format
9. Publish current reading value from the meter number 15 HEX in multipoint mode:
Command format
Response format
10.Read block of data from EEPROM in point-to-point mode:
Command format
Response format
11.Display "HELLO" onto the meter number 15 HEX in multipoint mode:
Command format
Response format
*15U01<CR>
15U01@<CR>
*E01<CR>
E01<CR>
*15D04<CR>
15D04<CR>
*Z04<CR>
Z04<CR>
*V01<CR>
V01 567.891 567.880 712.345 110.765<CR>
*15X01<CR>
X01 567.891<CR>
*R42<CR>
R42271100010001EO3E003F<CR>
*15Y01HELLO<CR>
15Y01<CR>
8-17