10. Data Format Commands (P, G, R, W)
10.13 COMMUNICATIONS BUS FORMAT ("bUS.FMt")
BIT POSITION
7
6
5
4
0
1
0
1
0
1
0
EXAMPLE: To configure communication protocol and bus format as following:
checksum excluded, no line feed, echo mode on, point to point, command mode and
RS-232 protocol:
The command data is 00010100 BIN. = 14 HEX. Then send *W1C14
TABLE 10.13 BUS FORMAT
3
2
1
0
0
1
0
1
0
1
0
1
bUS.FMt
Checksum - EXCLUDED
Checksum - INCLUDED
NO Line Feed
Line Feed following with <CR.
NO ECHO
ECHO
Point to Point Mode (RS-232)
Multi-Point (RS-485)
CONTINUOUS (Point to Point Mode)
COMMAND
MESSAGE (Continuous MODE)
CHARACTER (Continuous MODE)
RS-232
RS-485
ALWAYS 0, (Not Used)
10-13