Vo1" Response Data Format - Omega DP41-B Series User Manual

Universal input
Hide thumbs Also See for DP41-B Series:
Table of Contents

Advertisement

8. Command and Response Structure
8. Command and Response Structure
8. Command and Response Structure
8. Command and Response Structure
8. Command and Response Structure

8.7 "VO1" RESPONSE DATA FORMAT

As described in Section 8.3 response to the "V01" Command has the following format
if in echo mode:
[nn][V01][Sa[b]]S<value>[S<value>][S<value>][S<value>][<SP>uuu][hh]<CR>[<LF>]
"nnV01" is omitted from the response format if in no-echo mode; where:
CHARACTERS
a
b
first<value>
second<value>
third<value>
fourth<value>
<SP>
S
<uuu> 1
<uuu> 2
hh
<CR>
LF
[ ]
The length of the message transmitted by the meter in response to the "V01" command
is variable and controlled by the two menu items as "dAt FMt" (Data Format) and
"bUS.FMt" (Bus Format). (See Sections 10.12 and 10.13).
TABLE 8.6 SPECIAL CHARACTERS
Alarm status character
Peak/valley status character
Value of current reading
Value of filtered reading
Value of peak reading
Value of valley reading
Space
Space or <CR>
Three characters for the unit of measurement
N/A
Checksum
Carriage return
Line feed
Optional item
8-15

Advertisement

Table of Contents
loading

Table of Contents