Preparing the Thermostatted Column Compartment for
Transportation
When
If a Thermostatted Column Compartment including the Valve Drive Option shall be transported
Tools required
Screwdriver Pozidriv 1 PT3
Parts required
#
1
The module has been shipped with transportation locks, which must be used for
N O T E
transportation protection. Transportation Lock Kit (p/n G1316-67001) can be re-ordered.
1
Remove the valve head as described in
Heads"
on page 142.
1290 Infinity TCC User Manual

Preparing the Thermostatted Column Compartment for Transportation

p/n
Description
G1316-67001
Transportation Lock Kit
"Replacing Valve
2
Replace the valve head by the transportation valve head.
Bring the transportation valve head to the outer position.