Agilent Technologies 1290 Infinity User Manual page 44

Thermostatted column
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

2
Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see
place the module on almost any desk or laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for air circulation and electric connections.
If the bench should carry an Agilent system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Condensation
Condensation within the module
C A U T I O N
Condensation will damage the system electronics.
44
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Table 13
on page 45) allow you to
1290 Infinity TCC User Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents