Agilent Module to Agilent 35900 A/D Converters
p/n 5061-3378
Agilent Module to General Purpose
p/n 01046-60201
1290 Infinity II Preparative Open-Bed Fraction Collector User Manual
Pin 35900
Pin Agilent
A/D
module
1 - White
1 - White
2 - Brown
2 - Brown
3 - Gray
3 - Gray
4 - Blue
4 - Blue
5 - Pink
5 - Pink
6 - Yellow
6 - Yellow
7 - Red
7 - Red
8 - Green
8 - Green
9 - Black
9 - Black
Wire Color
Pin Agilent
module
White
1
Brown
2
Gray
3
Blue
4
Pink
5
Yellow
6
Red
7
Green
8
Black
9
Identifying Cables
Remote Cables
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not
connected
Power on
High
Ready
High
Stop
Low
Start request
Low
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not
connected
Power on
High
Ready
High
Stop
Low
Start request
Low
9