Electrical requirements
Voltage
100 - 240 Vac
Power
35 VA max.
Frequency
50/60 Hz (+/- 3Hz)
Fuse 800 mA/250 Vac, 5 mm x 20 mm, Slo-Blo (line and neutral fused)
Environmental
Protection (EN 60529)
Operating temperature
Storage temperature
Mechanical weight
Wiring the power connector (See Fig. II.1)
Hot leads: L and N
Ground:
Minimum diameter of power connection cable: 0.75 mm
Protective earthing
It is necessary to connect the protective conductor
terminal on the rear panel to the star point of machine
ground. Minimum cross-section of the connecting wire:
2
6 mm
, see Fig. II.2.
Preventative maintenance
No special preventative maintenance is necessary. For cleaning, wipe
lightly with a dry lint-free cloth.
POSITIP 880
IP 40 back panel
IP 54 front panel
0° to 45°C (32° to 113°F)
-20° to 70°C (-22° to 158°F)
3.2 kg (7.2 lb.)
Fig. II.1 Wiring the power connector
2
.
Fig. II.2 The protective conductor terminal on the
rear panel.