Analog Outputs - GE Optica Operator's Manual

General eastern dew point analyzer
Hide thumbs Also See for Optica:
Table of Contents

Advertisement

2.4.1 Analog Outputs

Note: When the Optica is being programmed, the analog outputs provide 4–20 mA and 0–5 VDC signals representing
the designated parameters.
For 4–20mA output, connect to terminals labelled 4–20 (+) and RTN (–).
Note: The maximum load allowed for current output is 500 Ohms.
For 0–5 VDC output, connect to terminals labelled 0–5 (+) and RTN (–).
Note: The maximum current allowed for voltage output is 5 mA.
Example:
Assume a temperature output, scaled to range from 0°C (T
of 23°C (T
):
actual
The voltage output is calculated by:
yielding an output voltage of 1.15V.
The current output is calculated by:
yielding an output current of 7.68 mA.
Optica™ Operator's Manual
(
T
actual
--------------------------------------- -
=
V
out
(
T
upper
(
)
23 0
×
-------------------- -
(
)
100 0
(
T
actual
(
)
--------------------------------------- -
I
mA
=
out
(
T
upper
(
)
23 0
-------------------- -
×
(
16
(
)
100 0
) to 100°C (T
), with a measured actual temperature
lower
upper
)
T
lower
×
5
)
T
lower
5
1.15V
=
)
T
lower
×
(
)
20 4
+
)
T
lower
)
7.68mA
4
+
=
Chapter 2. Installation
4
15