Chapter 2. Installation

2.4.3 Serial Output

The Serial Output connector is located at the lower left of the rear panel of the bench-mount unit, and the bottom panel
of the wall-mount unit. The output provides RS-232 serial communications between the unit and a terminal or a PC
running in terminal emulation mode.
The connector is a standard 9-pin D connector. For connection to another serial device, the cable is wired as shown in
Table 1 below. For a basic interface without handshaking, only pins 2, 3 and 5 (RX, TX and GND) on the Optica
connector are needed. Pin connections are given for both 25-pin and 9-pin devices.
The baud rate, format of the data, number of stop bits, number of data bits, and parity can all be programmed using the
menus.
18
Table 1: Serial Output Wiring Connections
Optica Connector
Pin
Connection
2
(RX)
3
(TX)
4
(DTE)
5
(GND)
6
(DSR)
7
(RTS)
8
(CTS)
25-Pin Device
Pin
Connection
3
(TX)
2
(RX)
6
(DSR)
7
(GND)
20
(DTE)
4
(CTS)
5
(RTS)
9-Pin Device
Pin
Connection
3
(TX)
2
(RX)
6
(DSR)
5
(GND)
4
(DTE)
8
(CTS)
7
(RTS)