7.4.4 "Balance" Remains on the Status Line

When the word
Balance
recovered from a
PACER
Check that the sensor and sensor cable are connected. If necessary, connect them, and the unit will complete
the
cycle after a short time (5 to 15 minutes).
PACER
The sensor optical bridge may be out of balance. Refer to "Cleaning and Balancing the Sensor Mirror" on
page 75 for instructions.

7.4.5 No Analog Output

If there is no analog output, but the digital display indicates correctly, check the analog output scaling. For the
see "Analog Outputs" on page 44. For the 4x40 unit, see Function 1 in "Analog Outputs" on page 62.

7.4.6 No Serial Output

Check the serial port programming for proper settings. For the
on page 55. For the
4x40
Optica™ Operator's Manual
remains displayed on the front panel for more than 15 minutes, the instrument has not
cycle.
unit, see the instructions in "Serial Output Setup" on page 63.
unit, see the instructions in "Serial Output Data"
VGA
Chapter 7. Maintenance
unit,
VGA
83