5:00 p.m. (Mountain Time), excluding all company and legal holidays. Telephone support is provided for troubleshooting and determination of parts to be shipped from GE Analytical Instruments to the customer in order to return the product to oper- ation. If telephone support is not effective, the product may be returned to GE Analytical Instruments for repair or replace- ment.
Page 16
Customers shall pay freight charges, including all taxes and duties, for all prod- ucts returning to GE Analytical Instruments. Any product returned to the factory without an RA number will be returned to the customer.
The Operation and Maintenance Manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present. In no event shall GE Analytical Instruments be liable for in- cidental or consequential damages in connection with or arising from the use of the manual and its accompanying related materials.
Les procédures de correction indiquées dans le présent document sont les seuls remèdes du client. Le groupe GE Analytical Instruments ne saurait en aucun cas être tenu pour responsable de tout préjudice direct, indirect ou spécial de quelque na- ture que ce soit (y compris, les pertes de bénéfices), qu’il soit fondé...
Gli unici rimedi spettanti all’utente sono quelli qui inclusi. In nessun caso GE Analytical Instruments sarà responsabile per dan- ni diretti, indiretti, speciali, accidentali o consequenziali (inclusa la perdita di profitti) risultanti dall’applicazione del contratto, atto illecito o altra teoria legale.
Page 23
This is a Safety Class I product. It must be must be attached to a grounded power source. Warning If this instrument is used in a manner not specified by GE Analytical Instruments USA, the protection provided by the instrument may be impaired. Warning Hazardous reagents (ammonium persulfate and phosphoric acid) are used in the Analyzer.
Page 24
Warning For performance within specifications on ozonated water systems, an Ozone Destruct Kit must be purchased from GE Analytical Instruments and installed according to instructions. Warning To avoid false TOC readings and possible damage to the Analyzer, always make sure the sample inlet is open and the DI water reservoir is filled before starting analysis.
Page 27
Este producto es de clase de seguridad I. Debe conectarse a una fuente de alimentación con toma de masa. Advertencia Si este instrumento se utiliza de una manera no especificada por GE Analytical Instruments USA, la protección ofrecida por el instrumento puede verse reducida. Advertencia El analizador utiliza reactivos peligrosos (persulfato de amonio y ácido fosfórico).
Page 28
Advertencia Para trabajar dentro de las especificaciones en sistemas de agua ozonizada, se deberá adquirir un kit de destrucción de ozono a GE Analytical Instruments e instalarlo de acuerdo con las instrucciones. Advertencia Para evitar falsas lecturas de TOC y posibles daños al analizador, asegúrese siempre de que la entrada de muestras está...
Page 31
à la terre. Avertissement Si cet instrument est utilisé de manière non conforme à ce qui est spécifié par le groupe GE Analytical Instruments aux USA, la protection fournie par l’instrument risque d’être réduite. Avertissement Des réactifs chimiques dangereux (persulfate d’ammonium et acide phosphorique) sont utilisés dans l’analyseur.
Page 32
Pour des performances optimales avec les spécifications des systèmes d’eau ozonée, un kit de destruction d’ozone (Ozone Destruct Kit) doit être obtenu auprès d’GE Analytical Instruments et installé conformément aux instructions données. Avertissement Afin d’éviter les mesures TOC erronées et d’éventuels dommages de l’analyseur, assurez-vous toujours que l’entrée de l’échantillon est ouverte et que le réservoir...
Page 35
Dies ist ein Produkt der Sicherheitsstufe I. Es muss an eine geerdete Stromquelle angeschlossen werden. Warnung Wenn dieses Instrument in einer Art und Weise verwendet wird, die nicht von GE Analytical Instruments USA festgelegt ist, kann der durch dieses Instrument gebotene Schutz beeinträchtigt werden.
Page 36
Maßnahmen ergreifen. Warnung Für die Funktion gemäß Spezifikationen für sauerstoffangereicherte Wassersysteme muss ein Ozone Destruct Kit für GE Analytical Instruments erworben werden und gemäß den Anweisungen installiert werden. Warnung Stellen Sie vor dem Beginn der Analyse stets sicher, dass der Probeneinlauf offen und der DI-Wasserbehälter gefüllt ist, um falsche Gesamtkohlenstoffmesswerte...
Page 39
Questo è un prodotto che rientra nella classe di sicurezza I. Deve essere collegato a una sorgente di alimentazione con messa a terra. Avvertenza Se lo strumento viene utilizzato in modo diverso da quello specificato da GE Analytical Instruments USA, la protezione fornita dallo strumento può risultare compromessa.
Page 40
Avvertenza Per un rendimento in linea con le specifiche su sistemi ad acqua ozonata, acquistare un Ozone Destruct Kit presso GE Analytical Instruments e installarlo nel modo indicato. Avvertenza Per evitare letture di TOC non corrette e possibili danni all’analizzatore, accertarsi sempre che l’ingresso del campione sia aperto e il serbatoio di acqua deionizzata...
Chapter 1. Introduction The Sievers™ 900 Series Total Organic Carbon Analyzers from GE Analytical Instruments includes high-sensitivity Analyzers used to measure the concentration of total organic carbon (TOC), total inorganic carbon (TIC), and total carbon (TC = TOC + TIC) in water samples (for patent information, see the section called “Trademarks and Patents”...
Analyzer design and function. An interior overview diagram is available in the “Maintenance” chapter on page 172. If you need additional assistance, contact GE Analytical Instruments Technical Support at 800.255.6964. Installation and training by a qualified service technician also can be provided.
Figure 4: Right Side Dimensions of the Analyzer Step 4: Install the Reagent Cartridges The oxidizer and acid reagents are shipped from GE Analytical Instruments in specific packaging for safety in transit. Carefully read the attached MSDS sheets prior to opening the packaging. Within the packaging are two reagent cartridges.
The Analyzer has a 25-pin parallel (Centronics) printer port for connecting an optional printer. The printer must be IBM compatible or Epson FX 850 compatible. GE Analytical Instruments sells a Seiko and a Citizen printer for use with the Analyzer.
Note that serial communication is supported only for computers running the Windows operating system. Note: For long distance connections (up to 1,000 feet or 305 meters), GE Analytical Instruments recommends use of either a RS-232 current loop converter or a RS-422 converter. Companies such as B B Electronics manufacture both kinds of converter.
Enabling DataGuard or Password Protection (Optional) The Sievers 900 Series of TOC Analyzers offers two levels of security, one included with all Analyzers and the other available for purchase as an upgrade from GE Analytical Instruments. The Password Protection feature is included with all Analyzers and provides a basic level of security.
An overview of the menu structure is given in “The Menu Screen” on page 91. An overview of hardware features starts on page 119. When the Analyzer is first powered on, the Startup screen displays basic GE Analytical Instruments contact information and the firmware version number. The Analyzer then begins initialization to check the status of consumables and prepare for taking measurements.
Setting Reagent Flow Rates GE Analytical Instruments recommends using the Sievers 900 On-Line TOC Analyzer in Autoreagent mode to ensure the optimum reagent flow rates. In this mode, the Analyzer will automatically calculate and implement the proper reagent flow rates. This feature greatly simplifies the process of manually determining optimum reagent flow rates.
Page 95
The sample flow rate increases when new tubing is installed, and the flow rate stabilizes after a few days of operation. For that reason, the tubing used in the Sievers 900 Series Analyzers has been "broken in" at the factory so that the flow rate is relatively stable immediately after it is installed.
For all samples, verify the pH of waste is <2 and adjust acid flow rate as necessary Understanding Analysis Times Analysis times in Sievers 900 Series TOC Analyzers vary depending on the type of reagent settings you use, either Autoreagent mode or manual reagents mode. Analysis times for measurements using the Autoreagent setting will be longer (and the volume of sample consumed is greater), as the Analyzer must identify the correct flow rates to use before starting full analysis.
Hyperterminal to communicate with the Analyzer, see “Using HyperTerminal” on page 228 in the Appendix. Note: For long distance connections (up to 1,000 feet or 305 meters) GE Analytical Instruments recommends use of either a RS-232 current loop converter or a RS-422 converter. Companies such as B&B Electronics manufacture both kinds of converter.
“Warnings and Error Messages” on page 196. Displaying System Information If you need to contact GE Analytical Instruments about your Analyzer, you can find most essential information on a single screen.
Chapter 5. Password Protection and DataGuard Overview The Sievers 900 Series of TOC Analyzers offers two levels of security, one included with all Analyzers and the other available as an optional upgrade from GE Analytical Instruments. The Password Protection feature is included with all Analyzers and provides a basic level of security.
DataGuard offers support for regulation 21 CFR Part 11, providing multiple levels of user access to the Analyzer via up to 100 unique User IDs and maintaining an audit trail of all user operations. DataGuard must be purchased as an upgrade from GE Analytical Instruments; once it is enabled, it cannot be disabled.
Once you inactivate the default administrator account, it cannot be used to access the Analyzer. If you forget the login information for the new administrator account that you have created, contact GE Analytical Instruments for assistance. Adding User IDs After DataGuard has been enabled, you can add a User ID by following these steps: Select the DataGuard tab.
The Model 900 TOC Analyzers can be calibrated using one of two methods: single-point calibration at 1, 5, 10, 25 or 50 ppm; or multi-point calibration at 1, 5, 10, 25 and 50 ppm. GE Analytical Instruments recommends that the Analyzer be calibrated in the field using a single-point calibration at a concentration that encompasses the range of interest.
The Sievers 900 Series TOC Analyzers facilitate the calibration and verification process by handling the necessary calculations internally; you will have the opportunity to accept or reject the calibration during the procedure. Calibration and verification are accomplished using Sievers Standards in 40-mL vials, sampled via the Analyzer’s IOS System.
TOC and IC in the selected concentration. For customers typically operating below 1 ppm, GE Analytical Instruments recommends selecting the 1 ppm single-point calibration. The available concentrations are shown in Table 12.
Figure 26: IC Values Summary Screen in a Multi-Point Calibration Performing a Calibration Verification Use the verification procedure to confirm that the Analyzer’s current calibration is accurate. GE Analytical Instruments recommends performing a verification after replacement of consumables, such as the pump tubing, UV lamp, resin bed, or reagents, and after calibration.
Performing an Autoreagent Verification Use the Autoreagent verification to confirm the Analyzer's current Autoreagent calibration is accurate. GE Analytical Instruments recommends performing the Autoreagents verification after replacement of pump tubing and after Autoreagent calibration. The sample flow rate calibration (see page 141) should always be performed after replacement of the pump tubing and before the Autoreagents verification.
Refer to Table 17 below for the recommended maintenance schedule. Chemical reagents, the in-line filter, UV lamp, and pump tubing must be purchased from GE Analytical Instruments. The use of reagents from other sources or failure to replace the reagents on the prescribed replacement schedule will invalidate the Analyzer’s warranty.
Table 17: Replacement Schedule for Sievers 900 Series Analyzer Consumables Description Typical Operating Life DI Water Reservoir Refill as needed (check when replacing reagents or UV lamp) Oxidizer 3 calendar months maximum (depends on flow rate) Acid 12 calendar months maximum (depends on flow rate)
Replacing the Chemical Reagents The oxidizer and acid reagents must be purchased from GE Analytical Instruments to ensure product purity and operating life. The use of reagents from other sources or failure to replace the reagents on the prescribed replacement schedule will invalidate the Analyzer’s warranty.
The tubing for the sample pump loses elasticity over time and must be replaced annually to ensure proper flow rates. The tubing must be purchased from GE Analytical Instruments; use of tubing from other sources or failure to replace the tubing on the prescribed replacement schedule will affect Analyzer functionality.
25. Enter the installation date for the pump tubing by following the procedures in “Setting the Installation or Expiration Date for New Consumables” on page 190. 26. GE Analytical Instruments recommends that you now perform a sample flow rate calibration, as described on page 141.
(prior to purification), the filter element will need to be replaced more often than if monitoring the water after purification. If the filter element clogs too frequently, contact GE Analytical Instruments to receive help in the selection of larger-capacity filters. As the filter is used, the flow rate of water through the IOS System will decrease and can even stop.
Note: the DI water reservoir should be checked and filled as necessary as part of every maintenance. A copy of this worksheet should be completed for each year the Analyzer is in service. The worksheet is available from the GE Analytical Instruments web site (www.GEInstruments.com).
This chapter provides the starting point for troubleshooting basic issues with your Analyzer. If you need additional assistance, in the U.S. contact GE Analytical Instruments Technical Support at 800.255.6964. In other parts of the world, contact your local GE Analytical Instruments representative.
A complete list of warnings and errors is shown below. Warnings and Error Messages 2 - Analog Warning Message: “An analog error was detected. Turn power off and on to reset. Contact GE if the error persists.” Explanation: This warning generally displays when the temperature or conductivity for the measurement is out of range.
Page 197
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the resin bed life remains. Order a new resin bed from GE now to avoid down time when the resin bed expires. 8 - User List Warning Message: “The user list is nearly full.
Page 198
Message: “The estimated acid life is less than 15 days. Order a new acid container.” Explanation: Based on the expiration date, the Analyzer estimates that only 15 days of acid life remain. Order a new acid container from GE now to avoid down time when the acid expires. 14 - Oxid Warning Message: “The estimated oxidizer life is less than 15 days.
Page 199
Refer to the 900 ICR Operation and Maintenance Manual for troubleshooting suggestions. 25 - History Data Warning Message: “A problem has been corrected in the history data. Contact GE if the warning persists.” Explanation: The Analyzer has detected a problem reading or writing to the data history memory;...
Page 200
Support to confirm that the Analyzer is functioning properly. 51 - History Flash Error Message: “A problem has been detected in the history data flash. Please contact GE.” Explanation: The Analyzer experienced a problem when trying to store data in the flash memory. It is possible a section of the flash memory is damaged.
If the error persists, contact Technical Support. 67 - Internal Error Message: “An internal error has been detected. Please contact GE.” Explanation: This error displays when a software error occurs in the Analyzer. Contact Technical Support.
While the Analyzer is operating, open the cover and make sure the sample pump is turning. If the pump is not turning and the Analyzer is operating (the display screen indicates that analysis is taking place), contact GE Analytical Instruments Technical Support for instructions. Warning To avoid potentially dangerous shock, do not touch anything inside the Analyzer while observing the sample pump.
To eliminate the problems encountered in TOC measurements of water samples containing high levels of IC, GE Analytical Instruments offers an inline ICR accessory. The module is installed on the Analyzer and uses a membrane-based system and vacuum to remove approximately 99% of the IC is at concentrations up to 25 ppm.
Step 4: Contact Technical Support After pursuing Steps 1 through 3, in the USA contact GE Analytical Instruments Technical Support at 800.255.6964. In other parts of the world, contact your local GE Analytical Instruments representative. When you call technical support, please have basic Analyzer information from the System Information screen (Maintenance tab System Info) available.
Page 215
Analyzer. Note that you may need an electrician to help you disconnect the Analyzer from the power supply. You can now repackage the Analyzer and ship it to GE Analytical Instruments. For international shipments, coordinate with a GE Analytical Instruments representative to ensure quick passage through customs.
Turbo sampling mode is an option on a Sievers 900 On-Line TOC Analyzer that is achieved by purchasing an activation code from GE Analytical Instruments. Turbo mode is useful for monitoring recycle or reclaim water in semiconductor facilities where rapid response is of primary concern.
Instrument Delay (90% recovery) 3.5 minutes Enabling Turbo Mode Activation of Turbo sampling mode may be purchased directly from GE Analytical Instruments. Select the Setup tab. Press the Enable Turbo button. Enter the Activation Code. If necessary, press the Number button to toggle the keypad from letters to numbers.
Turbo calibration, the IOS System is not used, and instead standards are sampled directly from the 250 mL containers provided by GE Analytical Instruments. In addition, only a single-point calibration is performed. The Analyzer must be in Turbo mode to perform the calibration.
Place the open end of the sipper tube into the bottle containing the TOC calibration standard (4 ppm KHP) that was purchased from GE Analytical Instruments, and press Next. 10. When prompted, remove the sipper tube from the TOC calibration standard bottle, place it into the bottle containing the IC calibration standard (as Na ), and press Next.